|
The UniLux is a single parameter fluorometer which can detect one of Chlorophyll a, Phycocyanin, Phycoerythrin for algae monitoring and Rhodamine or fluorescein for dye tracing applications. It is also available as a nephelometer for turbidity monitoring.
Available parameters:
|
- Applications and Features
Applications - In-situ chlorophyll & algae class studies
- Environment monitoring
- Dye tracing
- Particulate studies
- Process control
- Moored, profiled, towed or ROV / AUV platforms
Features
- Miniature, low cost single wavelength fluorometer
- Range of wavelength available
- Digital output in engineering units
- Additional analogue output as standard
- User selectable sampling rate, 0.1Hz - 3Hz
- Low power consumption
- Internal referencing of excitation intensity
- Rejection of ambient daylight
- Low noise, high sensitivity
- Low turbidity breakthrough
- User adjustable dynamic range
- RS232 & analogue output (or RS422 option)
- Description
The innovative UniLux range of digital in-situ fluorometers provides the user with increased functionality when compared to other fluorometers currently available on the market. CTG has used expertise gained within its Life Science Group that have resulted in a significant reduction in production costs, providing the marine, fresh water, surveying and process control markets with a low cost, high performance sensor. Users can select from a wide range of fluorescent parameters, specifying the UniLux tailored to their specific monitoring requirements.
Two signals can be obtained from the standard UniLux: the first is a digital RS232 output in engineering units, the second a 0V to 5V volt calibrated analogue output derived from the digital signal. A single RS422 output is also available as an option. This flexibility makes the UniLux suitable for integrating into many different systems and platforms.
A Windows based Graphical User Interface (GUI) is provided that allows the user to both plot and record time stamped data when operating the UniLux directly from a PC and gives control over many instrument parameters, including: sampling rate, calibration factors and dynamic range.
The dynamic range of the UniLux can be adjusted simply by varying the LED intensity using the GUI, while still maintaining the factory set calibration via the internal referencing of LED output.
Although small and low cost, the UniLux fluorometer maintains the performance standards and build quality previously seen in the CTG’s range of in-situ fluorometers.
Screen shot from the UniLux GUI application
- Specification
The UniLux is calibrated over the following ranges:
Performance
Wavelength Option
Dynamic Range*
Limit of Detection
Chlorophyll a
0 to 100 μg/L
<0.01 μg/L
Fluorescein
0 to 100 μg/L
<0.05 μg/L
Rhodamine WT
0 to 100 μg/L
<0.02 μg/L
Phycoerythrin Cyanobacteria
0 to 100 μg/L
<0.02 μg/L
Phycocyanin Cyanobacteria
0 to 100 μg/L
<0.01 μg/L
Turbidity
0 to 100 FTU
<0.02 FTU
*User configurable up to 500 μg/L
Mechanical
Size:
26.5mm dia x 105mm (140mm including connector)
Weight in air:
100g
Pressure housing:
Acetal C
Depth rating:
600 metres
Connector:
MCBH-6-MP-SS
Electrical
Input Voltage:
11 to 25Vdc
Data Output:
Digital RS232 and analogue 0 to 5Vdc (RS422 & SDI-12 options)
Power Requirements:
<1Watt @ 12 volt
Delivery Package: The standard UniLux fluorometer is supplied with a transit case, handbook and Windows based Graphical User Interface (GUI) Accessories: User to specify accessories. In view of our policy of continual improvement, the designs and specifications of our products may vary from those described
User to specify cables and other accessories.- Accessories
- The standard UniLux fluorometer is supplied with transit case, handbook and Windows based Graphical User Interface (GUI). There is now a range of accessories available to compliment the UniLux (see below). For full desciptions on each of the accessories, please refer to datasheet in download section or visit our shop. You will need to decide on cable type, length and cable termination. You will also need to give consideration to how you wish to interface the fluorometer to either a laptop, PDA or your own system.If you need advice regarding which accessories you may need, please contact This e-mail address is being protected from spambots. You need JavaScript enabled to view it Tel: +44(0)20 8481 9019.

- Unilux Downloads